Off-resonance-robust velocity-selective magnetization preparation for non-contrast-enhanced peripheral MR angiography.
Abstract
PURPOSE: To develop a new velocity-selective (VS) excitation pulse sequence which is robust to field inhomogeneity, and demonstrate its application to non-contrast-enhanced peripheral MR angiography (MRA).
METHODS: The off-resonance-robust VS saturation pulse is designed by incorporating 180° refocusing pulses into the k-space-based reference design and tailoring sequence parameters in a velocity region of interest. The VS saturation pulse is used as magnetization preparation for non-contrast-enhanced peripheral MRA to suppress background tissues but not arterial blood based on their velocities. Non-contrast-enhanced peripheral MRA using the proposed VS preparation was tested in healthy volunteers and a patient with arterial stenosis.
RESULTS: Calf angiograms obtained using the new VS preparation show more uniform background suppression than the reference VS preparation, as demonstrated by larger mean values and smaller standard deviations of artery-to-vein and artery-to-muscle contrast-to-noise ratios (71.0 ± 11.4 and 75.3 ± 12.1 versus 61.7 ± 22.7 and 58.5 ± 27.8). Two-station peripheral MRA using the new VS preparation identifies stenosis of the femoral and popliteal arteries in the patient, as validated by digital subtraction angiography.
CONCLUSION: Non-contrast-enhanced MRA using the new VS magnetization preparation can reliably provide high angiographic contrast in the lower extremities with significantly improved immunity to field inhomogeneity.
Research topics
Related publications
View all publications- Luo J, Addy NO, Ingle RR, Hargreaves BA, Hu BS, Nishimura DG, Shin T.Magn Reson Med. 74(6):1632-9. doi: 10.1002/mrm.25575. Epub 2014 Dec 17.2015 Dec 01
- Malavé MO, Baron CA, Addy NO, Cheng JY, Yang PC, Hu BS, Nishimura DG.Magn Reson Med. doi: 10.1002/mrm.27475. [Epub ahead of print]2018 Oct 29
- Addy NO, Ingle RR, Wu HH, Hu BS, Nishimura DG.Magn Reson Med. 74(3):614-21. doi: 10.1002/mrm.25803. Epub 2015 Jul 14.2015 Sep 01
- Luo J, Addy NO, Ingle RR, Baron CA, Cheng JY, Hu BS, Nishimura DG.Magn Reson Med. doi: 10.1002/mrm.26273. [Epub ahead of print]2016 May 13
- Publication
Nonrigid autofocus motion correction for coronary MR angiography with a 3D cones trajectory.
Ingle RR, Wu HH, Addy NO, Cheng JY, Yang PC, Hu BS, Nishimura DG.Magn Reson Med. 72(2):347-61. doi: 10.1002/mrm.24924. Epub 2013 Sep 4.2014 Aug 01