Non-contrast-enhanced peripheral angiography using a sliding interleaved cylinder acquisition.
Abstract
PURPOSE: To develop a new sequence for non-contrast-enhanced peripheral angiography using a sliding interleaved cylinder (SLINCYL) acquisition.
METHODS: A venous saturation pulse was incorporated into a three-dimensional magnetization-prepared balanced steady-state free precession sequence for non-contrast-enhanced peripheral angiography to improve artery-vein contrast. The SLINCYL acquisition, which consists of a series of overlapped thin slabs for volumetric coverage similar to the original sliding interleaved ky (SLINKY) acquisition, was used to evenly distribute the venous-suppression effects over the field of view. In addition, the thin-slab-scan nature of SLINCYL and the centric-ordered sampling geometry of its readout trajectory were exploited to implement efficient fluid-suppression and parallel imaging schemes. The sequence was tested in healthy subjects and a patient.
RESULTS: Compared to a multiple overlapped thin slab acquisition, both SLINKY and SLINCYL suppressed the venetian blind artifacts and provided similar artery-vein contrast. However, SLINCYL achieved this with shorter scan times and less noticeable artifacts from k-space amplitude modulation than SLINKY. The fluid-suppression and parallel imaging schemes were also validated. A patient study using the SLINCYL-based sequence well identified stenoses at the superficial femoral arteries, which were also confirmed with digital subtraction angiography.
CONCLUSION: Non-contrast-enhanced angiography using SLINCYL can provide angiograms with improved artery-vein contrast in the lower extremities.
Research topics
Related publications
View all publications- Shin T, Hu BS, Nishimura DG.Magn Reson Med. 70(5):1229-40. doi: 10.1002/mrm.24561. Epub 2012 Nov 28.2013 Nov 01
- Wang TY, Xiao L, Alexander KP, Rao SV, Kosiborod MN, Rumsfeld JS, Spertus JA, Peterson ED.Circulation. 118(21):2139-45. doi: 10.1161/CIRCULATIONAHA.108.787143. Epub 2008 Nov 3.2008 Nov 18
- Malavé MO, Baron CA, Addy NO, Cheng JY, Yang PC, Hu BS, Nishimura DG.Magn Reson Med. doi: 10.1002/mrm.27475. [Epub ahead of print]2018 Oct 29
- Publication
Nonrigid autofocus motion correction for coronary MR angiography with a 3D cones trajectory.
Ingle RR, Wu HH, Addy NO, Cheng JY, Yang PC, Hu BS, Nishimura DG.Magn Reson Med. 72(2):347-61. doi: 10.1002/mrm.24924. Epub 2013 Sep 4.2014 Aug 01 - Osnabrugge RL, Arnold SV, Reynolds MR, Magnuson EA, Wang K, Gaudiani VA, Stoler RC, Burdon TA, Kleiman N, Reardon MJ, Adams DH, Popma JJ, Cohen DJ; CoreValve U.S. Trial Investigators.JACC Cardiovasc Interv. 8(2):315-23. doi: 10.1016/j.jcin.2014.08.016.2015 Feb 01